FDA Recall Terminated

Surgical microscope The device is used to improve the visibility of objects through magnification and illumination and can be applied for observation and documentation and for human veterinary medical treatment.

Recall: Z-2176-2013 · Initiated December 18, 2012

Recall

Recall Number
Z-2176-2013
Event Number
64601
Firm
Leica Microsystems (Schweiz) Ag 201 Max Schmidheiny-Strasse Heerbrugg Switzerland
FEI Number
3003974370
Product Code
FSO
Status
Terminated
Root Cause
Device Design
Initiated
December 18, 2012
Posted
September 10, 2013
Terminated
April 24, 2017

Description

Surgical microscope The device is used to improve the visibility of objects through magnification and illumination and can be applied for observation and documentation and for human veterinary medical treatment.

Reason

The brake in the Z axis and/or brakes in the AB and C axes failed during installation.

Action

The firm, Leica Microsystems, sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 19, 2012 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to DISCONTINUE USE until your system software is updated and to complete and return the attached Acknowledgement Form via email at [email protected] or fax to: 847-236-3747 or +41 71 726 32 49 within 10 working days after receipt of field safety notice. Your Leica Mycrosystems will contact you to make the necessary arrangements to complete an upgrade at no charge to the customers. If you have any further questions, contact Regulatory Affairs/Quality Assurance Manager at 847-405-5413 or +41 71 726 32 16.

Distribution

Worldwide distribution: USA (Nationwide) including states of: AK, CA, CO, NY, and OR; and countries of: Belarus, China, Hong Kong, India, Indonesia, Italy, Norway, South Africa, Spain, Sweden, Turkey, and Vietnam.

Quantity

5 units