FDA Recall Terminated

Brilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728240. Computed Tomography X-ray system

Recall: Z-2175-2018 · Initiated March 30, 2018

Recall

Recall Number
Z-2175-2018
Event Number
80094
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Device Design
Initiated
March 30, 2018
Terminated
April 14, 2020
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Brilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728240. Computed Tomography X-ray system

Reason

The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.

Action

Urgent Field Safety Notices dated 3/28/18 were distributed to customers. The letters instructed customers to perform the following actions: Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please retain a copy with the equipment Instruction for Use. As outlined in the Instructions for Use, always monitor the patient during all movements (manual or motorized) of the patient support. If you recognize an impending collision with the back shelter wall, halt the motion or activate the system E-stop. Mandatory Field Change Order (FCO) 72800686 which will add additional warning labels specific to this issue to the system. The FCO will be implemented free of charge.

Distribution

Worldwide distribution including US states of CA, MD, NV, OH, TX, and WI, Afghanistan, Canada, Germany, Iraq, Italy, Korea, Kuwait, and Lebanon.

Quantity

2