FDA Recall Terminated

Tool, Stripper W/BLD, 12S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.

Recall: Z-2174-2015 · Initiated May 28, 2015

Recall

Recall Number
Z-2174-2015
Event Number
71534
Firm
American Medical Systems Innovation Center Silicon Valley
FEI Number
2937094
Product Code
GEX
Status
Terminated
Root Cause
Reprocessing Controls
Initiated
May 28, 2015
Posted
July 21, 2015
Terminated
January 19, 2016
Address
3070 Orchard Dr, San Jose, CA, 95134-2011

Description

Tool, Stripper W/BLD, 12S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.

Reason

validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products

Action

Urgent Medical Device Field notification letters were sent on June 9, 2015. Translation for countries out of the US will be conducted. Communications are being coordinated by the Prostate Health QA Director. Urgent Medical Device Field notification Letters advised customers of the issue and list all affected products with name and model numbers. Any SureFlex Reuseable Fibers which have not been used but are in inventory and with the 3 year shelf life should be returned as per instructions enclosed. Any of the SureFlex devices in inventory that have been used should be discarded or returned. All Reuseable Stripper and Cleaver Accessories that have not been used should be returned. Used devices should be destroyed or returned. Questions to 1-408-456-3320. The Acknowledgement form should be completed and returned.

Distribution

Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.

Quantity

16,710 total affected devices