FDA Recall Open, Classified

Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Recall: Z-2173-2024 · Initiated May 13, 2024

Recall

Recall Number
Z-2173-2024
Event Number
94678
Firm
SIGHT DIAGNOSTICS LTD 23, Derekh Menakhem Begin Tel Aviv-Yafo Israel
FEI Number
3016099075
Product Code
GKZ
Status
Open, Classified
Root Cause
Process control
Initiated
May 13, 2024
Posted
June 21, 2024

Description

Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Reason

One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.

Action

On May 15, 2024, URGENT Medical Device Recall letters were sent to customers. Following identification of the affected lot units, please take the following action: 1. Please discontinue use and quarantine any product from Lot 10114. 2. Return any unused units of test kit Lot 10114 to the company or discard them immediately according to standard practice. NOTE: For return of unused test kits please indicate so in the acknowledgement and receipt form and Customer Support will contact you with further instructions. 3. Review previously obtained test results conducted with the affected lot and identify populations at higher risk. Retest if necessary. NOTE: If you require assistance in identifying samples tested using the affected lot please indicate so in the acknowledgment and receipt form and we will be happy to provide you with any necessary information. 4. If you are a distributor, please share this notification with any customer to whom the product was shipped, or may have been shipped. 5. You are kindly requested to fill out the attached Acknowledgement and Receipt Form and return it to Sight as soon as possible, and in any case no later than 30 days from the date of this notification, via email, to [email protected]. Please provide the form even if you no longer have any product in inventory so that we can verify that all affected product has been removed. 6. If you are required to notify your local authority please use this notification and notify Sight as soon as possible. We appreciate your time and apologize for any inconvenience this may have caused. Please do not hesitate to reach out with any questions or feedback to Customer Support via the following channels: US Customers [email protected] Callback service (voicemail) +1-833-SIGHTDX (7444839 ext. 3) Text +1 617 4105226 Partners (Distributors) [email protected] Callback service (voicemail) +44 0203 093 0772 (ext. 2) Text +44 7449 063724 Customer Support is available Mon-Fri EST- 2:00AM-7:00

Distribution

Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.

Quantity

7,450 units (2,800 US, 4,650 OUS)