FDA Recall Terminated

Elecsys PreciControl Varia 3

Recall: Z-2173-2017 · Initiated April 10, 2017

Recall

Recall Number
Z-2173-2017
Event Number
77108
Firm
Roche Diagnostics Corporation
FEI Number
1823260
Product Code
JJY
Status
Terminated
Root Cause
Process control
Initiated
April 10, 2017
Posted
May 31, 2017
Terminated
August 4, 2017
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Elecsys PreciControl Varia 3

Reason

e-barcode and value sheet for PreciControl Varia contains a typographical error for the target values for the Elecsys PTH (parathyroid hormone) and Elecsys PTH STAT immunoassays

Action

On 4/10/2017, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected consignees via courier. The recall notifications included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.

Distribution

Nationwide

Quantity

1603