FDA Recall Terminated

KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 US highway 202 South Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 USA The cobas HPV Test is a qualitative in vitro test for the detection of Human Papillomavirus in patient specimens. The test utilizes amplification of target DNA by the Polymerase Chain Reaction (PCR) and nucleic acid hybridization for the detection of 14 high-risk (HR) HPV types in a single analysis. The test specifically identifies types HPV16 and HPV18 while concurrently detecting the rest of the high risk types (31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66 and 68). The cobas HPV Test is indicated: (a) To screen patients 21 years and older with ASC-US (Atypical squamous cells of undetermined significance) cervical cytology test results to determine the need for referral to colposcopy. (b) To be used in patients 21 years and older with ASC-US cervical cytology results, to assess the presence or absence of high-risk HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. The results of this test are not intended to prevent women from proceeding to colposcopy. (c) In women 30 years and older, the cobas HPV Test can be used with cervical cytology to adjunctively screen to assess the presence or absence of high risk HPV types. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. (d) In women 30 years and older, the cobas HPV Test can be used to assess the presence or absence of HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. Cervical specimens that may be tested with the cobas HPV Test include the following liquid based collection media and collection device: " ThinPrep Pap TestTM PreservCyt Solution " Endocervical Brush/Spatula

Recall: Z-2173-2013 · Initiated July 25, 2013

Recall

Recall Number
Z-2173-2013
Event Number
66036
Firm
Roche Molecular Systems, Inc.
FEI Number
2243471
Product Code
MAQ
Status
Terminated
Root Cause
Device Design
Initiated
July 25, 2013
Posted
September 10, 2013
Terminated
August 6, 2015
Address
1080 Us Highway 202 S, Branchburg, NJ, 08876-3733

Description

KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 US highway 202 South Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 USA The cobas HPV Test is a qualitative in vitro test for the detection of Human Papillomavirus in patient specimens. The test utilizes amplification of target DNA by the Polymerase Chain Reaction (PCR) and nucleic acid hybridization for the detection of 14 high-risk (HR) HPV types in a single analysis. The test specifically identifies types HPV16 and HPV18 while concurrently detecting the rest of the high risk types (31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66 and 68). The cobas HPV Test is indicated: (a) To screen patients 21 years and older with ASC-US (Atypical squamous cells of undetermined significance) cervical cytology test results to determine the need for referral to colposcopy. (b) To be used in patients 21 years and older with ASC-US cervical cytology results, to assess the presence or absence of high-risk HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. The results of this test are not intended to prevent women from proceeding to colposcopy. (c) In women 30 years and older, the cobas HPV Test can be used with cervical cytology to adjunctively screen to assess the presence or absence of high risk HPV types. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. (d) In women 30 years and older, the cobas HPV Test can be used to assess the presence or absence of HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. Cervical specimens that may be tested with the cobas HPV Test include the following liquid based collection media and collection device: " ThinPrep Pap TestTM PreservCyt Solution " Endocervical Brush/Spatula

Reason

During an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas 4800 HPV Master Mix reagent that was packaged in four (4) cobas 4800 AMP/DET kits, US- and CE-IVD (240 and 960 test kit configurations). Examination of the batch production record of the material revealed there was an error in the tare weight used for the carboy during the fina

Action

Roche Molecular Diagnostics sent an Urgent Medical Device Correction letter dated July 25, 2013, to all affected customers via UPS return receipt. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to: 1. Immediately discontinue use of the product lots listed above. 2. Discard affected product according to local guidelines. 3. To receive replacement inventory, complete the entire attached faxback form and fax it to 1-866-918-8741. 4. The faxback form must be completed and returned even if you do not have affected inventory. 5. File this Urgent Medical Device Correction (UMDC) for future reference. Please contact Roche Molecular Diagnostics Technical Support, 24 hours a day, seven days a week at 1-800-526-1247, if you have questions about the information contained in this UMDC.

Distribution

Worldwide Distribution - USA (nationwide) including the countries of Austria, Australia, Belgium, Brazil, Czech Rep, Canada, Croatia, Columbia, Denmark, Dominican Republic, Germany, Ecuador, France, Japan, Greenland, Mexico, Italy, New Zealand, Netherlands, Peru, Norway, Singapore, Romania, South Africa, Spain, South Korea, Sweden, Switzerland and UK.

Quantity

US-IVD: 350 Kits; CE-IVD: 441 KITS