FDA Recall Open, Classified

GlideScope Go Monitors

Recall: Z-2172-2021 · Initiated July 1, 2021

Recall

Recall Number
Z-2172-2021
Event Number
88214
Firm
Verathon, Inc.
FEI Number
3022472
Product Code
CCW
Status
Open, Classified
Root Cause
Device Design
Initiated
July 1, 2021
Address
20001 N Creek Pkwy, Bothell, WA, 98011-8218

Description

GlideScope Go Monitors

Reason

Handheld video monitor failed to meet the labeled IP67 rating which may allow fluid ingress into the monitor resulting in loss of functionality.

Action

On 07/01/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via FedEx to customers informing them that the Recalling Firm has become aware of a variation in material thickness of a micro USB mounting plate can lead to insufficient sealing, resulting in the device failing to meet the labeled IP67 rating allowing fluid ingress and impact functionality of the monitor. Customers are instructed to: 1. Perform a visual inspection of monitor for fluid ingress, referring to figures 1 and 2 of the Notification Letter for visual examples of the failure to meet the IP67 rating. 2. Based on the visual inspection of their device(s), do one of the following: -If their Monitor has no fluid ingress or they do not wish to return their affected product, complete the attached Recall Response Form by indicating no or 0 in the text box and return the form by email to [email protected]. Or -If their Monitor has fluid ingress and they would like these units serviced (similar to Figures 1 and 2) complete the attached Recall Response Form. They will need to indicate, by number, the count of Monitors that require service in the text box and return the form by email to [email protected]. The Recalling Firm's Customer Care will contact them to make arrangements to replace their monitor. The firm is implementing a change in the design for the monitor micro-USB mounting plate to reduce the variation in the thickness in order to meet the IP67 rating. Any questions - contact the Recalling Firm's Representative or Customer Care at [email protected]

Distribution

U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Australia, Brunei Darussalam, Canada, Czech Republic, Denmark, Estonia, Falkland Island (Malvinas), France, Germany, Greece, Hong Kong, Iceland, Ireland, Israel, Italy, Kuwait, Luxembourg, Malaysia, New Zealand, Oman, Papua New Guinea, Philippines, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, United Arab Emerita, and United Kingdom.

Quantity

8,569