FDA Recall Terminated

Architect Sirolimus Reagent Pack; 100 test pack containing 1 bottle (8.0 mL) Anti-Sirolimus Coated Microparticles, 1 bottle (8.0 mL) Sirolimus Acridinium-labeled Conjugate in Citrate Buffer, and 1 bottle (10 mL) Assay Diluent containing saline; Abbott Laboratories, Abbott Diagnostics Division, Abbott Park, IL 60064 USA; list 01L76-25. The Architect Sirolimus assay is chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sirolimus in human whole blood on the Architect i System, as an aid in the management of renal transplant patients receiving sirolimus therapy.

Recall: Z-2172-2010 · Initiated May 27, 2010

Recall

Recall Number
Z-2172-2010
Event Number
55871
Firm
Abbott Laboratories
FEI Number
1415939
Product Code
NRP
Status
Terminated
Root Cause
Equipment maintenance
Initiated
May 27, 2010
Posted
August 3, 2010
Terminated
July 15, 2011
Address
100 Abbott Park Rd, Abbott Park, IL, 60064-3502

Description

Architect Sirolimus Reagent Pack; 100 test pack containing 1 bottle (8.0 mL) Anti-Sirolimus Coated Microparticles, 1 bottle (8.0 mL) Sirolimus Acridinium-labeled Conjugate in Citrate Buffer, and 1 bottle (10 mL) Assay Diluent containing saline; Abbott Laboratories, Abbott Diagnostics Division, Abbott Park, IL 60064 USA; list 01L76-25. The Architect Sirolimus assay is chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sirolimus in human whole blood on the Architect i System, as an aid in the management of renal transplant patients receiving sirolimus therapy.

Reason

The Architect Sirolimus assay lot 80162M100 is exhibiting higher than normal reports of barcode read errors.

Action

Abbott sent a "Product Recall" letter dated May 27, 2010, to all customers receiving the affected lot of reagent. The customers were informed of the increased frequency of barcode read errors due to the print quality of the barcode label. The following error codes have been observed for some kits within the lot: 4000 - Unable to read reagent bar code in position (x) on (y) carousel, 0208 - Reagent carousel scan error, and 2011 - Extra reagent bottle detected in position (x) on (y) carousel. The customers were requested to determine if they are currently using or have inventory of lot 80162M100, discontinue use of and destroy any remaining inventory of the lot, and retain a copy of the letter for their laboratory records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688. Customers outside the U.S., please contact your local Customer Service representative.

Distribution

Worldwide distribution: USA including Puerto Rico, and countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Israel, Italy, Japan, Latvia, Lithuania, Malaysia, Mexico, New Zealand, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, Venezuela and Vietnam.

Quantity

875 kits