FDA Recall Open, Classified

Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;

Recall: Z-2169-2025 · Initiated June 18, 2025

Recall

Recall Number
Z-2169-2025
Event Number
97130
Firm
Medtronic, Inc.
FEI Number
2182208
Product Code
NVZ
Status
Open, Classified
Root Cause
Software design
Initiated
June 18, 2025
Posted
July 23, 2025
Address
8200 Coral Sea St Ne, Mounds View, MN, 55112-4391

Description

Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;

Reason

In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.

Action

An URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letter dated 6/18/25 was sent to customers. Customer Actions: "Update SmartSync (to version 4.2.3 or higher). Refer to release notes provided by your local Medtronicrepresentative for update instructions. If needed, Medtronic representatives are available to work with you toinstall or update the SmartSync application(s) on your tablet. See Appendix for instructions on how to verify thesoftware update is complete. "Sign and return the enclosed confirmation form and share this notice with those who need to be aware withinyour organization or with any organization where SmartSync may be in use. Questions regarding this informationshould be directed to the local Medtronic Representative or Medtronic Programmer Technical Services at 800-638-1991, Option 1.

Distribution

Worldwide - US Nationwide distribution.

Quantity

22091 total devices