FDA Recall Terminated

Statspin Express 4 Centrifuge Model Number: SSH4 Intended Use: Rapid Separation of plasma and serum from primary gel collection tubes

Recall: Z-2169-2011 · Initiated August 18, 2010

Recall

Recall Number
Z-2169-2011
Event Number
57446
Firm
StatSpin, Inc., dba Iris Sample Processing
FEI Number
1221015
Product Code
JQC
Status
Terminated
Root Cause
Device Design
Initiated
August 18, 2010
Posted
May 10, 2011
Terminated
September 1, 2020
Address
60 Glacier Dr, Westwood, MA, 02090-1800

Description

Statspin Express 4 Centrifuge Model Number: SSH4 Intended Use: Rapid Separation of plasma and serum from primary gel collection tubes

Reason

Rotors may fail due to wire spokes breaking

Action

Iris notified customers by letter in August 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were requested to install the new rotors and discard the older rotors with the wires. There was no need for customers to return the rotors, but the firm requested customers fax or e-mail the enclosed form to verify that they have installed the new rotors. For any questions customers should call 1-800-782-8774.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Canada, Australia Belgium, Czech Republic, Finland, France, Greece, Kuwait, Norway, Turkey,UK, and Norway.

Quantity

2090 units