Statspin Express 4 Centrifuge Model Number: SSH4 Intended Use: Rapid Separation of plasma and serum from primary gel collection tubes
Recall
- Recall Number
- Z-2169-2011
- Event Number
- 57446
- Firm
- StatSpin, Inc., dba Iris Sample Processing
- FEI Number
- 1221015
- Product Code
- JQC
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- August 18, 2010
- Posted
- May 10, 2011
- Terminated
- September 1, 2020
- Address
- 60 Glacier Dr, Westwood, MA, 02090-1800
Description
Statspin Express 4 Centrifuge Model Number: SSH4 Intended Use: Rapid Separation of plasma and serum from primary gel collection tubes
Rotors may fail due to wire spokes breaking
Iris notified customers by letter in August 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were requested to install the new rotors and discard the older rotors with the wires. There was no need for customers to return the rotors, but the firm requested customers fax or e-mail the enclosed form to verify that they have installed the new rotors. For any questions customers should call 1-800-782-8774.
Worldwide Distribution - USA (nationwide) and the countries of Canada, Australia Belgium, Czech Republic, Finland, France, Greece, Kuwait, Norway, Turkey,UK, and Norway.
2090 units