FDA Recall Terminated

Suction Jar Lid for Medela Vario 18 Vacuum Pump used with the Bravo pH Monitoring System Medtronic Part Number: 9043K2101. (9043K2111 is the OUS version) The Vacuum Pump is used to provide the vacuum pressure during Capsule attachment. Correct Overflow lid: Medela CH-6341 Baar Switzerland Art. No. 077.0440, Small lid with conical patient connection 6-10mm and overflow protection device.(Incorrect overflow lid is Medela P/N: 077.0450)

Recall: Z-2168-2008 · Initiated April 25, 2008

Recall

Recall Number
Z-2168-2008
Event Number
48612
Firm
Medtronic Neuromodulation
FEI Number
2182207
Product Code
BTA
Status
Terminated
Root Cause
Employee error
Initiated
April 25, 2008
Posted
September 4, 2008
Terminated
May 15, 2012
Address
800 53rd Ave NE PO Box 1250, Minneapolis, MN, 55440-1250

Description

Suction Jar Lid for Medela Vario 18 Vacuum Pump used with the Bravo pH Monitoring System Medtronic Part Number: 9043K2101. (9043K2111 is the OUS version) The Vacuum Pump is used to provide the vacuum pressure during Capsule attachment. Correct Overflow lid: Medela CH-6341 Baar Switzerland Art. No. 077.0440, Small lid with conical patient connection 6-10mm and overflow protection device.(Incorrect overflow lid is Medela P/N: 077.0450)

Reason

The Medela Vario 18 Vacuum Pump used with the Medtronic Bravo pH Monitoring System. The Vacuum Pump is used to provide the vacuum pressure during Capsule attachment. Some Medela Vario 18 Vacuum Pumps were provided with an incorrect suction Jar Lid that did not have the correct conical connection for the Vacuum Tube.

Action

Affected customers contacted by letter "Medtronic Urgent: Medical device Recall April 2008" that indicates that some Medela Vario 18 Vacuum Pumps were provided with an incorrect Suction Jar Lid that did not have the correct Conical Connection for the Vacuum Tube and the method to obtain a replacement part if needed. The letter requires customers to locate the Vacuum Pump serial number and compare it against the Serial Number on the Reply Form provided. Next customers are to review the Suction Jar Lid and check the part number. "Reply form and Instruction for Receiving Replacement Lid " included. Contact Medtronic 1-800-707-0933 for assistance.

Distribution

USA including states of OH, DE, CT, KY, AR, NY, MA, CA, GA, NC, MI, WI, TX, PA, MD, and country of CANADA.

Quantity

1500 pumps (20 affected)