FDA Recall Open, Classified

Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Recall: Z-2167-2025 · Initiated June 26, 2025

Recall

Recall Number
Z-2167-2025
Event Number
97222
Firm
Hamilton Medical AG Parc Industrial Vial 10 Domat/Ems Switzerland
FEI Number
3014766741
Product Code
BZO
Status
Open, Classified
Root Cause
Process control
Initiated
June 26, 2025
Posted
July 30, 2025

Description

Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Reason

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.

Action

Starting on 6/26/2025, recall notices were emailed to customers who were asked to do the following: Return affected devices. If one of the affected breathing circuits would be already in use with a patient, it is requested to exchange it and discard the product after replacement. If you are a distributor, please forward to any organization where potentially affected product has been transferred. Complete and return the acknowledgement form via email to [email protected] If you have any questions contact firm at: [email protected] and 800-426-6331

Distribution

US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.

Quantity

2,560