FDA Recall Terminated

Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product is manufactured and distributed by Xlumena, Inc., Mountain View, CA The NAVIX Access Device is a multi-lumen balloon catheter with a trocar, anchor balloon, and dilation balloon that creates and dilates a tract across tissue planes. The NAVIX Access Device is indicated for use as an accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst, when it is visibly bulging into the gastrointestinal tract during endoscopic procedures.

Recall: Z-2167-2013 · Initiated April 17, 2013

Recall

Recall Number
Z-2167-2013
Event Number
64965
Firm
Xlumena, Inc.
FEI Number
3008516478
Product Code
GCJ
Status
Terminated
Root Cause
Process design
Initiated
April 17, 2013
Posted
September 9, 2013
Terminated
September 23, 2013
Address
453 Ravendale Dr, Ste H, Mountain View, CA, 94043-5221

Description

Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product is manufactured and distributed by Xlumena, Inc., Mountain View, CA The NAVIX Access Device is a multi-lumen balloon catheter with a trocar, anchor balloon, and dilation balloon that creates and dilates a tract across tissue planes. The NAVIX Access Device is indicated for use as an accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst, when it is visibly bulging into the gastrointestinal tract during endoscopic procedures.

Reason

A complaint investigation has found that product fractures can occur at the distal end of the catheter under load.

Action

Xlumena sent an Urgent Medical Device Recall letter dated April 17, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and segregate the affected product. Customers with questions were instructed to contact [email protected]. For question regarding this recall call 650-961-9901.

Distribution

Worldwide Distribution - USA including AZ, CA, FL, GA, IN, NY,SC, WA, PA, IL, CO, and internationally to Netherlands, Germany, Spain, Denmark, and Switzerland.

Quantity

148 units