FDA Recall Terminated

SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.

Recall: Z-2166-2017 · Initiated January 31, 2017

Recall

Recall Number
Z-2166-2017
Event Number
76770
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
CDZ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 31, 2017
Posted
April 19, 2017
Terminated
May 24, 2018
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.

Reason

Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.

Action

Siemens issued an Urgent Medical Device Correction (UMDC) to customers using the affected SHBG adjustor lots. Customers were advised to store the affected reconstituted adjustor lots at 2-8C or -20C for up to 14 days, after which they must be discarded. The IMMULITE/IMMULITE 1000 SHBG is not affected by this issue. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding this recall call 914-631-8000.

Distribution

Nationwide Distribution

Quantity

9100 kits