FDA Recall Open, Classified

BD COR System Software. Model Number: 444829.

Recall: Z-2162-2025 · Initiated June 23, 2025

Recall

Recall Number
Z-2162-2025
Event Number
97142
Firm
Becton Dickinson & Co.
FEI Number
1119779
Product Code
MAQ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
June 23, 2025
Posted
July 23, 2025
Address
7 Loveton Cir, Sparks, MD, 21152-9212

Description

BD COR System Software. Model Number: 444829.

Reason

Potential for functionality issue that supports the Over labeling feature and could lead to possible false negative results.

Action

Becton, Dickinson and Company began notification of consignees on about 06/23/2025 via telephone call and followed by a letter. Consignees were instructed to note the outlined workflows when loading specimens with the Over labeling feature enabled and consider retesting affected specimens for HPV testing if affected. Consignees were also instructed to post the notification within their facility, notify customers if affected units were further distributed, and complete and return the provided Customer Response Form.

Distribution

Worldwide - US Nationwide distribution in the states of Florida, Texas and Michigan. The countries of Sweden and Belgium.

Quantity

7 systems