FDA Recall Terminated

CORFLO, PEG Adaptor Repair Kit with ENFit Connectors, Size: 16FR; Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG adapter is the connection point between the tubing and the feeding set/syringe. The adapter is designed as a replaceable component.

Recall: Z-2162-2017 · Initiated February 3, 2017

Recall

Recall Number
Z-2162-2017
Event Number
76465
Firm
CORPAK MedSystems, Inc.
FEI Number
3011270181
Product Code
KNT
Status
Terminated
Root Cause
Device Design
Initiated
February 3, 2017
Terminated
August 5, 2019
Address
1001 Asbury Dr, Buffalo Grove, IL, 60089-4528

Description

CORFLO, PEG Adaptor Repair Kit with ENFit Connectors, Size: 16FR; Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG adapter is the connection point between the tubing and the feeding set/syringe. The adapter is designed as a replaceable component.

Reason

Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.

Action

Firm provided consignees with field correction on 2/2/2017 by letter mailed via FedEx. Within the notice, the firm provided additional instructions for handling the PEG Y-Adapter to help reduce the likelihood for exposing the stem to excess stress that could break the connector. Consignees were instructed to forward the field correction notice to all staff members or end-users who use the ENFit Y-Adapter.

Distribution

Worldwide distribution in the states of NH, VT, CA and the countries of UK, Ireland, France, Austria, Norway, Sweden, New Zealand

Quantity

2 cases (5 units per case)