FDA Recall Terminated

Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100

Recall: Z-2160-2019 · Initiated June 18, 2019

Recall

Recall Number
Z-2160-2019
Event Number
83345
Firm
Howmedica Osteonics Corp.
FEI Number
3004024955
Product Code
NKB
Status
Terminated
Root Cause
Labeling mix-ups
Initiated
June 18, 2019
Terminated
May 22, 2020
Address
2 Pearl Ct, Allendale, NJ, 07401-1611

Description

Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100

Reason

Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.

Action

1. Please note this product has already been removed by your local Stryker Sales Representative and/or Stryker Branch or Agency. This communication is being made for notification purposes only; there are no additional product return requirements for this matter. 2. Please assist us in meeting our regulatory obligation by completing and signing the enclosed Product Recall Acknowledgment Form and email a copy to Stryker Spine at [email protected] or fax the form to 201-575-4675. If you have any questions, please contact Christa Joisil or Matthew Kelleher from the Regulatory Compliance Team: [email protected] or 201.749.8090

Distribution

US Nationwide Distribution - CA, FL, GA, MA, NY, PA, TX

Quantity

11