FDA Recall
Terminated
General X-Ray System types with Digital Spot Imaging (DSI) software
Recall: Z-2160-2008
·
Initiated May 23, 2008
Recall
- Recall Number
- Z-2160-2008
- Event Number
- 48602
- Firm
- Philips Medical Systems North America Co. Phillips
- FEI Number
- 1000524572
- Product Code
- IZJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- May 23, 2008
- Posted
- August 25, 2008
- Terminated
- September 28, 2010
- Address
- 22100 Bothell Everett Hwy, Bothell, WA, 98021-8431
Description
General X-Ray System types with Digital Spot Imaging (DSI) software
Reason
X-ray images may be stored in the wrong patient file or corrupted, images from two different patients may be combined in one patient's folder or may be combined on one image (corrupted).
Action
On 5/20/08 the firm issued an URGENT - DEVICE CORRECTION letter to the consignees advising of the problem, requesting consignees to check the stored or printed images to ensure that it represents the data and images captured from the examination. The firm will issue a Field Change Order to their service representatives providing new software, when software is available. This should be issued by the end of June.
Distribution
Nationwide Distribution.
Quantity
41 units