FDA Recall Terminated

DRG Salivary Testosterone ELISA- IVD for the detection of Testerone in human saliva Catalog # SLV-3013 Product Usage: The Salivary Testosterone ELISA (SlV-3013) is an enzymatic in vitro Assay for the detection of Testosterone in human Saliva. Testosterone levels are measured in fertility studies in men and women. Fertility problems are not diagnosed on the outcome of Testosterone alone but accompanied by other diagnostic means

Recall: Z-2159-2019 · Initiated May 21, 2019

Recall

Recall Number
Z-2159-2019
Event Number
83183
Firm
DRG Instruments GmbH Frauenbergstr. 18 Marburg Germany
FEI Number
3002800697
Product Code
CDZ
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 21, 2019
Terminated
May 29, 2020

Description

DRG Salivary Testosterone ELISA- IVD for the detection of Testerone in human saliva Catalog # SLV-3013 Product Usage: The Salivary Testosterone ELISA (SlV-3013) is an enzymatic in vitro Assay for the detection of Testosterone in human Saliva. Testosterone levels are measured in fertility studies in men and women. Fertility problems are not diagnosed on the outcome of Testosterone alone but accompanied by other diagnostic means

Reason

Salivary assay may observe a higher percentage of samples with 0 pg/ml testosterone or results which are too low. Saliva samples in the low measuring range of the kit(< 50 pg/ml) were assayed significantly lower

Action

DRG International issued notification via email on 5/21/19 advising of the problem, health risk and action to take: 1. We kindly ask you to immediately examine your stock and promptly quarantine these products, 2. Please send back or discard all unused kits. 3. Please use the attached response form, fill, sign and send back to us.4. Please inform your customers and forward this letter and notice to persons who received the affected lots and request they fill out the form and return at once.5. Patient samples that have been determined with the affected lots should be checked again carefully. If results are lower than expected samples should be re-run again. 6. Please tell us if you have experienced any problem with this lot and specify the details. Reply to DRG (email: [email protected] and copy [email protected]).

Distribution

Worldwide Distribution - US Nationwide CA, MI Foreign: Canada, Brazil, China, Poland

Quantity

38 kits