I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or warfarin.

Recall: Z-2157-2012 · Initiated March 1, 2012

Recall

Recall Number
Z-2157-2012
Event Number
62536
Firm
Abbott Point Of Care Inc.
FEI Number
2245578
Product Code
GJS
Status
Terminated
Root Cause
Process control
Initiated
March 1, 2012
Posted
August 8, 2012
Terminated
February 8, 2013
Address
400 College Rd E, Princeton, NJ, 08540-6607

Description

I-STAT PT/INR cartridge Abbott Point of Care Inc. Abbott Park, IL 60064 USA This cartridge is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or warfarin.

Reason

The i-STAT PT/INR cartridges have the potential to exhibit incorrectly elevated results.

Action

Abbott Point of Care Inc, sent a URGENT RECALL NOTICE dated March 2012 , to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were provided in the letter with instructions in the letter on how to handle the affected product. Please complete and return the enclosed Business Reply card to acknowledge receipt of this communication and installation of the interium software CLEW and also 1) Update all handhelds 2) download appropriate assignment sheets. For further questions please call (609) 454-9000.

Distribution

Worldwide Distribution -- (USA) nationwide

Quantity

2,950,296 cartridges US; 156,408 cartridges OUS