FDA Recall Terminated

Philips MultiDiagnost Eleva with Flat Detector (MD Eleva FD)Model: 708034 Product Usage: Philips X-ray System

Recall: Z-2156-2012 · Initiated July 11, 2012

Recall

Recall Number
Z-2156-2012
Event Number
62551
Firm
Philips Healthcare Inc.
FEI Number
1218950
Product Code
KPR
Status
Terminated
Root Cause
Software design
Initiated
July 11, 2012
Posted
August 8, 2012
Terminated
March 26, 2013
Address
3000 Minuteman Road, Andover, MA, 01810

Description

Philips MultiDiagnost Eleva with Flat Detector (MD Eleva FD)Model: 708034 Product Usage: Philips X-ray System

Reason

When doing a 3 Dimensional Rotational Angiography (3DRA) the 3D reconstruction does not always start, even when all images have been acquired.

Action

Philips issued an Urgent Field Safety Notice dated July 2, 2012 to all affected customers. The letter identified the affected product, problem description, actions to be taken by the customer/user in order to prevent risks for patients or users and actions planned by Philips to correct the problem. For question or support concerning this issue contact your local Philips representative.

Distribution

Worldwide Distribution - US (nationwide) including the states of CA, CT, MI, MN, MN, NC, NY, NY, VA and the countries of Belgium, Columbia, Czech Republic and Denmark. France Germany Italy Latvia Luxembourg Netherlands Russian Fed. Spain Sweden Switzerland Thailand United Kingdom

Quantity

41 units