Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General Surgical and Plastic Surgery applications for transection, osteotomy, graft harvesting, bone shaping, laminectomy, entry hole, suture hole, and midface advancement.
Recall
- Recall Number
- Z-2156-2008
- Event Number
- 48472
- Firm
- Medtronic Powered Surgical Solutions
- FEI Number
- 1625507
- Product Code
- HBB
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- May 30, 2008
- Posted
- August 23, 2008
- Terminated
- August 22, 2009
- Address
- 4620 N Beach St, Fort Worth, TX, 76137-3219
Description
Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General Surgical and Plastic Surgery applications for transection, osteotomy, graft harvesting, bone shaping, laminectomy, entry hole, suture hole, and midface advancement.
Cutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for revision surgery.
Firm notified consignees by Urgent: Medical Device Recall letter sent May 30, 2008. The letter requires consignees to immediately compare their inventory of products against the list provided and quarantine any unused dissecting tools. The letter also states that Medtronic would provide replacement product to match the quantity and type returned by the consignees' institution.
Nationwide Distribution.
1,306 units