FDA Recall Terminated

3.5mm X 32.5mm Cortical Screw used with the Polarus Humeral Fixation System,. Product labeled in part, "Polarus Humeral Fixation System& SIZE: 3.5mm X 32.5mm Cortical Screw... REF HCO3325-S... Acumed 5885 NW Cornelius Pass Road Hillsboro, OR 97124-9432 ". Polarus Humeral Fixation System, 3.5mm X 32.5mm Cortical Screw are packaged in a sterile inner package which are placed into an outer box. The cortical screws are used with the Polarus Locking Humeral Rod (Polarus Huemral Fixation System) for the multi-planar fixation. The screws are sold individually to support the Polarus Locking Humeral Rod systems in the field.

Recall: Z-2155-2008 · Initiated December 27, 2005

Recall

Recall Number
Z-2155-2008
Event Number
48468
Firm
Acumed LLC
FEI Number
1000125930
Product Code
HWC
Status
Terminated
Root Cause
Process change control
Initiated
December 27, 2005
Posted
August 28, 2008
Terminated
February 3, 2010
Address
5885 NW Cornelius Pass Rd, Hillsboro, OR, 97124-9370

Description

3.5mm X 32.5mm Cortical Screw used with the Polarus Humeral Fixation System,. Product labeled in part, "Polarus Humeral Fixation System& SIZE: 3.5mm X 32.5mm Cortical Screw... REF HCO3325-S... Acumed 5885 NW Cornelius Pass Road Hillsboro, OR 97124-9432 ". Polarus Humeral Fixation System, 3.5mm X 32.5mm Cortical Screw are packaged in a sterile inner package which are placed into an outer box. The cortical screws are used with the Polarus Locking Humeral Rod (Polarus Huemral Fixation System) for the multi-planar fixation. The screws are sold individually to support the Polarus Locking Humeral Rod systems in the field.

Reason

Package of 3.5mm X 32.5mm cortical screw may contain 3.5mm X 22.5mm cortical screw.

Action

The firm issued a Product Removal notification on 12-27-2005 to their customers. The notification was sent by either FAX or by e-mail. The Product Removal Notification informed their customers of the issue, asked them to remove the screws from inventory and return the screws to Acumed. A reminder notification, dated 1/20/08 was issued to customers identifying the product and the problem and requesting the product be returned.

Distribution

Worldwide Distribution --- USA including states of CA, KY, FL, TX, OH, OR, NY, ID NJ, NE, LA, WI, UT, MO, MS, GA, AZ and countries of United Kingdom, Italy, Puerto Rico, Turkey, Spain, Japan, South Africa, and Belgium.

Quantity

112 pieces