Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that will eventually resorb and potentially facilitate normalization of vessel function in patients with ischemic heart disease due to de novo native coronary artery lesions. The treated lesion length should be less than the nominal scaffolding length (8 mm, 12 mm, 18 mm, 23 mm, 28 mm) with reference vessel diameters e 2.0 mm and d 3.8 mm.
Recall
- Recall Number
- Z-2151-2017
- Event Number
- 77131
- Firm
- Abbott Vascular
- FEI Number
- 2024168
- Product Code
- PNY
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 26, 2017
- Posted
- June 1, 2017
- Terminated
- November 16, 2017
- Address
- 26531 Ynez Rd, Temecula, CA, 92591-4630
Description
Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that will eventually resorb and potentially facilitate normalization of vessel function in patients with ischemic heart disease due to de novo native coronary artery lesions. The treated lesion length should be less than the nominal scaffolding length (8 mm, 12 mm, 18 mm, 23 mm, 28 mm) with reference vessel diameters e 2.0 mm and d 3.8 mm.
Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.
Abbott Vascular sent an Urgent Medical Device Recall and Hazard Alert letter dated April 26, 2017, to affected customers to inform them that the action is in response to recent concerns over data from some studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent. Customers were informed of the actions to be taken. Customers were instructed to stop using these devices immediately, advise BVS patients to follow the recommendations for DAPT prescribed by their health care provider. Advise patients experiencing new cardiac symptoms to seek clinical care, review their inventory and complete the attached Facsimile Reply Form and display and share this notification with other relevant personnel within their organization. Customers with questions are instructed to contact their local Abbott Vascular Representative or Customer Service department.
Australia
590 units