NovaPACS versions 7.4, 7.5, 7.6, and 8.0, Diagnostic Viewer. Distributed by Novarad Corporation. Picture archiving and communication system (PACS)
Recall: Z-2151-2016
·
Initiated May 9, 2013
Recall
- Recall Number
- Z-2151-2016
- Event Number
- 74556
- Firm
- Novarad Corporation
- FEI Number
- 1000117423
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- May 9, 2013
- Terminated
- December 1, 2016
- Address
- 752 E 1180 S, Ste 200, American Fork, UT, 84003-3561
Description
NovaPACS versions 7.4, 7.5, 7.6, and 8.0, Diagnostic Viewer. Distributed by Novarad Corporation. Picture archiving and communication system (PACS)
Reason
Potential for images to be flipped while streaming, which could incorrectly display image orientation markers.
Action
Consignees were notified on 05/09/13 via letter and email to upgrade software and to turn off streaming function.
Distribution
Worldwide distribution including US nationwide. No Canadian distribution.
Quantity
2695