FDA Recall Terminated

NovaPACS versions 7.4, 7.5, 7.6, and 8.0, Diagnostic Viewer. Distributed by Novarad Corporation. Picture archiving and communication system (PACS)

Recall: Z-2151-2016 · Initiated May 9, 2013

Recall

Recall Number
Z-2151-2016
Event Number
74556
Firm
Novarad Corporation
FEI Number
1000117423
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
May 9, 2013
Terminated
December 1, 2016
Address
752 E 1180 S, Ste 200, American Fork, UT, 84003-3561

Description

NovaPACS versions 7.4, 7.5, 7.6, and 8.0, Diagnostic Viewer. Distributed by Novarad Corporation. Picture archiving and communication system (PACS)

Reason

Potential for images to be flipped while streaming, which could incorrectly display image orientation markers.

Action

Consignees were notified on 05/09/13 via letter and email to upgrade software and to turn off streaming function.

Distribution

Worldwide distribution including US nationwide. No Canadian distribution.

Quantity

2695