FDA Recall Open, Classified

TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 (20-foot)10-, 15- and 20-foot cords respectively), used with the Tablo Hemodialysis System.

Recall: Z-2150-2024 · Initiated April 30, 2024

Recall

Recall Number
Z-2150-2024
Event Number
94654
Firm
Outset Medical, Inc.
FEI Number
3010355846
Product Code
KDI
Status
Open, Classified
Root Cause
Device Design
Initiated
April 30, 2024
Posted
June 14, 2024
Address
3052 Orchard Dr, San Jose, CA, 95134-2011

Description

TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 (20-foot)10-, 15- and 20-foot cords respectively), used with the Tablo Hemodialysis System.

Reason

A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.

Action

On 4/30/24, correction notices were sent to customers asking them to do the following: The firm has enclosed the replacement power cord (PN-0008460) with this communication based on the number of affected devices at your facility. Disseminate this information to appropriate personnel within your facility for processing and distribution. This notice should be shared with any organization where the potentially affected devices have been transferred. Follow instructions to replace the power cords. Dispose of the affected power cords. Complete and return the acknowledgement and receipt form via email to [email protected] If you have questions or notice any damage around the power cord, please contact us at [email protected] or 844-MYTABLO (844-698-2256) and request an inspection from a Field Service Engineer.

Distribution

US Nationwide distribution in the states of FL, CA, NC, GA, MD, DE, AK, OK, MA, WV, IN, KY, AZ, LA, NE, KS, NJ, CT, NM, CO, SC, TX, PA, NY, OH.

Quantity

431