FDA Recall Terminated

SIEMENS ADVIA Centaur CP Immunoassay System BNP kit; SMN 10309045 - 500 test For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur CP System.

Recall: Z-2150-2014 · Initiated June 23, 2014

Recall

Recall Number
Z-2150-2014
Event Number
68665
Firm
Siemens Healthcare Diagnostics, Inc
FEI Number
1219913
Product Code
NBC
Status
Terminated
Root Cause
Other
Initiated
June 23, 2014
Posted
June 26, 2015
Terminated
March 31, 2016
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

SIEMENS ADVIA Centaur CP Immunoassay System BNP kit; SMN 10309045 - 500 test For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur CP System.

Reason

Complaints of failed calibration.

Action

Siemens sent an Urgent Medical Device Correction (UMDC) letter to all affected Siemens Healthcare Diagnostics customers in the United States on June 23, 2014 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on June 23, 2014 via hard copy, email, and/or fax per each country's local regulations. These notices inform the customer of the potential for low calibrator deviation failures when using ADVIA Centaur Systems BNP kit lots ending in 172 on the ADVIA Centaur CP due to an error on the master curve card. Siemens informs customers if a successful calibration is achieved, the results obtained are valid. If customers are unable to obtain a successful calibration on the ADVIA Centaur CP, they should switch to an alternative lot of BNP assay. The notices should be reviewed with the facility's laboratory or medical director. Fax-back forms will be used to confirm that the customers received the UMDC. For those customers that do not return a fax-back form, a Siemens representative will follow up with those customers to ensure they received the Urgent Medical Device Correction and completed the required actions. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding this recall call 877-229-3711.

Distribution

Worldwide Distribution - USA (nationwide including Puerto Rico) and Internationally to Canada, Albania, Argentina, Australia, Bosnia Herzeg., Brazil, Chile, Colombia, Egypt, Georgia, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Japan, Kazakhstan, Lebanon, Malaysia, New Caledonia, Oman, P.R. China, Pakistan, Philippines, Qatar, Rep. of Yemen, Republic Korea, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, Taiwan, Thailand, U.A.E., Ukraine, Venezuela, Vietnam, Austria, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece,Hungary,Iceland, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Poland, Portugal, Romania,Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Quantity

961 kits