FDA Recall Terminated

Chromovision High Definition (HD) Wireless Camera, Pt # CC-9715900. Visual recording of surgical procedures.

Recall: Z-2148-2012 · Initiated July 10, 2012

Recall

Recall Number
Z-2148-2012
Event Number
62683
Firm
Berchtold Corp.
FEI Number
1220685
Product Code
FWH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 10, 2012
Posted
August 6, 2012
Terminated
June 6, 2013
Address
1950 Hanahan Rd N, Charleston, SC, 29406-4878

Description

Chromovision High Definition (HD) Wireless Camera, Pt # CC-9715900. Visual recording of surgical procedures.

Reason

Defective camera: There is a possibility that the connection of the wires to the connector may become loose, causing image distortion or loss.

Action

BERCHTOLD Corporation sent a letter dated July 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. BERCHTOLD will replace the customer's cameras at no cost to the customer. Customers with questions should call 843-569-6100, ext 322. For questions regarding this recall call 843-569-6100.

Distribution

Nationwide Distribution including FL, MI, MS, and NH.

Quantity

17 cameras