FDA Recall Open, Classified

1. 3 VALVE MAN KIT, Medline Kit SKU Number: 65101544; 2. 3V MANIFOLD KIT, Medline Kit SKU Number: 65110351; 3. HEART CATH PACK-LF, Medline Kit SKU Number: DYNJ0373285M; 4. ANGIO DRAPE CARDIAC PACK, Medline Kit SKU Number: DYNJ51615P; 5. ANGIO DRAPE CARDIAC PACK, Medline Kit SKU Number: DYNJ51615R; 6. HEART RADIAL LAB PACK, Medline Kit SKU Number: DYNJ61982C; 7. HEART RADIAL LAB PACK, Medline Kit SKU Number: DYNJ61982D; 8. HEART ABLATION DRAPE PACK, Medline Kit SKU Number: DYNJ61984C; 9. TAVR PACK, Medline Kit SKU Number: DYNJ62310B; 10. HEART TAVR PACK, Medline Kit SKU Number: DYNJ65174G; 11. FEMORAL CARDIAC CATH, Medline Kit SKU Number: DYNJ66343A; 12. CARDIAC CATH PACK, Medline Kit SKU Number: DYNJ66693C; 13. TAVR PACK, Medline Kit SKU Number: DYNJ67225C; 14. AFIB TRAY, Medline Kit SKU Number: DYNJ85069; 15. AFIB TRAY, Medline Kit SKU Number: DYNJ85069A; 16. AFIB TRAY, Medline Kit SKU Number: DYNJ85069B; 17. MTO,LEFT HEART,KIT, Medline Kit SKU Number: VASC1811. Namic medical convenience kits, containing Namic Manifold, Labeled as: 1. DBD-MTS,3V,MAN,KIT, Medline Kit Number/SKU 60010195; 2. MTS,4 VALVE,LFT,HRT,KIT, Medline Kit Number/SKU 60010468; 3. MTO,ADD,ON,KIT, Medline Kit Number/SKU 60010469; 4. MTS,RT,HANDED/LFT,HRT,KIT, Medline Kit Number/SKU 60010508; 5. MTS,LFT,HRT,KIT, Medline Kit Number/SKU 60012911; 6. MTO,RIGHT HEART,KIT, Medline Kit Number/SKU 60012912; 7. DBD-MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60013001; 8. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60013101; 9. MTS,LFT,HRT,W/WIRE, Medline Kit Number/SKU 60013112; 10. MTS,-,3V,KIT, Medline Kit Number/SKU 60020608; 11. MTO,ANGIO,KIT, Medline Kit Number/SKU 60032681; 12. MTS-LEFT HEART,KIT, Medline Kit Number/SKU 60033123; 13. MTS,RIGHT HEART,KIT, Medline Kit Number/SKU 60033124; 14. MTO,LEFT HEART,KIT, Medline Kit Number/SKU 60034644; 15. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60040273; 16. MTO,2V,LEFT HEART,KIT, Medline Kit Number/SKU 60040424; 17. MTS,3,VAVLE,KIT, Medline Kit Number/SKU 60041082; 18. MTS,2V,KIT, Medline Kit Number/SKU 60042453; 19. MTO,3V,KIT, Medline Kit Number/SKU 60042454; 20. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60042721; 21. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60050183; 22. MTS,3 VALVE,MANIFOLD,KIT, Medline Kit Number/SKU 60060268; 23. MTO,RT,HRT,KIT, Medline Kit Number/SKU 60060564; 24. MTO,LEFT HEART,KIT, Medline Kit Number/SKU 60060569; 25. MTS, Medline Kit Number/SKU 60062021; 26. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60071232; 27. MTO,CATH,LAB,KIT, Medline Kit Number/SKU 60080221; 28. MTO,RIGHT, Medline Kit Number/SKU 60080347; 29. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60080349; 30. MTS,MANIFOLD,KIT, Medline Kit Number/SKU 60080716; 31. MTS,LEFT,HRT,KIT, Medline Kit Number/SKU 60081001; 32. MTS,RIGHT HEART,KIT, Medline Kit Number/SKU 60081002; 33. MTO,LEFT,HRT(RT,HANDED), Medline Kit Number/SKU 60081168; 34. MTO,RIGHT HEART,KIT, Medline Kit Number/SKU 60090123; 35. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60091111; 36. MTS,EP,KIT, Medline Kit Number/SKU 60100056; 37. MTS,3 VALVE,LEFT HEART,KIT, Medline Kit Number/SKU 60100338; 38. MTO,RIGHT HEART,KIT, Medline Kit Number/SKU 60100339; 39. DBD-MTO,LEFT HEART,KIT, Medline Kit Number/SKU 60110273; 40. MTO,RIGHT HEART,KIT, Medline Kit Number/SKU 60110701; 41. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60120693; 42. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60121817; 43. MTO,CATH,KIT, Medline Kit Number/SKU 60130124; 44. MTO,LEFT HEART,KIT, Medline Kit Number/SKU 60130318; 45. DBD-MTS,3 VALVE,KIT, Medline Kit Number/SKU 60130506; 46. MTO,RT HEART,KIT, Medline Kit Number/SKU 60131323; 47. MTS,LEFT,HRT,KIT, Medline Kit Number/SKU 60131324; 48. DBD-MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60131554; 49. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60131815; 50. MTO,RT,HRT,KIT, Medline Kit Number/SKU 60

Recall: Z-2145-2026 · Initiated March 24, 2026

Recall

Recall Number
Z-2145-2026
Event Number
98735
FEI Number
1417592
Product Code
OEZ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 24, 2026
Posted
May 8, 2026
Address
3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753

Description

1. 3 VALVE MAN KIT, Medline Kit SKU Number: 65101544; 2. 3V MANIFOLD KIT, Medline Kit SKU Number: 65110351; 3. HEART CATH PACK-LF, Medline Kit SKU Number: DYNJ0373285M; 4. ANGIO DRAPE CARDIAC PACK, Medline Kit SKU Number: DYNJ51615P; 5. ANGIO DRAPE CARDIAC PACK, Medline Kit SKU Number: DYNJ51615R; 6. HEART RADIAL LAB PACK, Medline Kit SKU Number: DYNJ61982C; 7. HEART RADIAL LAB PACK, Medline Kit SKU Number: DYNJ61982D; 8. HEART ABLATION DRAPE PACK, Medline Kit SKU Number: DYNJ61984C; 9. TAVR PACK, Medline Kit SKU Number: DYNJ62310B; 10. HEART TAVR PACK, Medline Kit SKU Number: DYNJ65174G; 11. FEMORAL CARDIAC CATH, Medline Kit SKU Number: DYNJ66343A; 12. CARDIAC CATH PACK, Medline Kit SKU Number: DYNJ66693C; 13. TAVR PACK, Medline Kit SKU Number: DYNJ67225C; 14. AFIB TRAY, Medline Kit SKU Number: DYNJ85069; 15. AFIB TRAY, Medline Kit SKU Number: DYNJ85069A; 16. AFIB TRAY, Medline Kit SKU Number: DYNJ85069B; 17. MTO,LEFT HEART,KIT, Medline Kit SKU Number: VASC1811. Namic medical convenience kits, containing Namic Manifold, Labeled as: 1. DBD-MTS,3V,MAN,KIT, Medline Kit Number/SKU 60010195; 2. MTS,4 VALVE,LFT,HRT,KIT, Medline Kit Number/SKU 60010468; 3. MTO,ADD,ON,KIT, Medline Kit Number/SKU 60010469; 4. MTS,RT,HANDED/LFT,HRT,KIT, Medline Kit Number/SKU 60010508; 5. MTS,LFT,HRT,KIT, Medline Kit Number/SKU 60012911; 6. MTO,RIGHT HEART,KIT, Medline Kit Number/SKU 60012912; 7. DBD-MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60013001; 8. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60013101; 9. MTS,LFT,HRT,W/WIRE, Medline Kit Number/SKU 60013112; 10. MTS,-,3V,KIT, Medline Kit Number/SKU 60020608; 11. MTO,ANGIO,KIT, Medline Kit Number/SKU 60032681; 12. MTS-LEFT HEART,KIT, Medline Kit Number/SKU 60033123; 13. MTS,RIGHT HEART,KIT, Medline Kit Number/SKU 60033124; 14. MTO,LEFT HEART,KIT, Medline Kit Number/SKU 60034644; 15. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60040273; 16. MTO,2V,LEFT HEART,KIT, Medline Kit Number/SKU 60040424; 17. MTS,3,VAVLE,KIT, Medline Kit Number/SKU 60041082; 18. MTS,2V,KIT, Medline Kit Number/SKU 60042453; 19. MTO,3V,KIT, Medline Kit Number/SKU 60042454; 20. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60042721; 21. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60050183; 22. MTS,3 VALVE,MANIFOLD,KIT, Medline Kit Number/SKU 60060268; 23. MTO,RT,HRT,KIT, Medline Kit Number/SKU 60060564; 24. MTO,LEFT HEART,KIT, Medline Kit Number/SKU 60060569; 25. MTS, Medline Kit Number/SKU 60062021; 26. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60071232; 27. MTO,CATH,LAB,KIT, Medline Kit Number/SKU 60080221; 28. MTO,RIGHT, Medline Kit Number/SKU 60080347; 29. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60080349; 30. MTS,MANIFOLD,KIT, Medline Kit Number/SKU 60080716; 31. MTS,LEFT,HRT,KIT, Medline Kit Number/SKU 60081001; 32. MTS,RIGHT HEART,KIT, Medline Kit Number/SKU 60081002; 33. MTO,LEFT,HRT(RT,HANDED), Medline Kit Number/SKU 60081168; 34. MTO,RIGHT HEART,KIT, Medline Kit Number/SKU 60090123; 35. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60091111; 36. MTS,EP,KIT, Medline Kit Number/SKU 60100056; 37. MTS,3 VALVE,LEFT HEART,KIT, Medline Kit Number/SKU 60100338; 38. MTO,RIGHT HEART,KIT, Medline Kit Number/SKU 60100339; 39. DBD-MTO,LEFT HEART,KIT, Medline Kit Number/SKU 60110273; 40. MTO,RIGHT HEART,KIT, Medline Kit Number/SKU 60110701; 41. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60120693; 42. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60121817; 43. MTO,CATH,KIT, Medline Kit Number/SKU 60130124; 44. MTO,LEFT HEART,KIT, Medline Kit Number/SKU 60130318; 45. DBD-MTS,3 VALVE,KIT, Medline Kit Number/SKU 60130506; 46. MTO,RT HEART,KIT, Medline Kit Number/SKU 60131323; 47. MTS,LEFT,HRT,KIT, Medline Kit Number/SKU 60131324; 48. DBD-MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60131554; 49. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60131815; 50. MTO,RT,HRT,KIT, Medline Kit Number/SKU 60

Reason

Medline has identified the presence of particulate within the fluid path of the Manifolds.

Action

Medline Industries, LP issued an URGENT MEDICAL DEVICE notice to its consignees on 3/24/2026 via email and mail. The notice explained the issue, potential risk and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-046 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or [email protected].

Distribution

US (nationwide) and AE, AU, CA, JP, KR, NL, SG, SK.

Quantity

567675 kits