FDA Recall Open, Classified

Beckman Coulter DxA Automation System, DxA 5000, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnly

Recall: Z-2144-2023 · Initiated June 2, 2023

Recall

Recall Number
Z-2144-2023
Event Number
92484
Firm
Beckman Coulter Biomedical GmbH Sauerbruchstr. 50 Munchen Germany
FEI Number
3006677029
Product Code
JJE
Status
Open, Classified
Root Cause
Software design
Initiated
June 2, 2023
Posted
July 14, 2023

Description

Beckman Coulter DxA Automation System, DxA 5000, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnly

Reason

A software defect in the DxA Automation System may cause tests to be run on patient samples in tube types not compatible with the requested test, leading to erroneous patient results

Action

On 06/12/2023, Beckman Coulter issued an Urgent Medical Device Recall Notification to affected consignees. Beckman Coulter asked consignees to take the following actions: 1. Inspect tubes routed to error regions and confirm that the tube type was appropriate per the assay IFU for the test performed. 2. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. 3. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email./" Manually, complete and return the enclosed Response Form. 4. If you have any questions regarding this notice, please contact our Customer Support Center or your local Beckman Coulter Representative: " From our website: http://www.beckmancoulter.com/" By phone: call 1-800-526-3821 in the United States. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Distribution

US: AL CA FL IL IN KY MI MN MO NE NJ NV NY OH PA SC TX WI WV OUS: Australia, Belgium, Canada, China, Croatia, Czechia, Egypt, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Saudi Arabia, Spain, Switzerland, United Kingdom,

Quantity

146 systems