FDA Recall Terminated

Aptio Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Recall: Z-2142-2020 · Initiated April 7, 2020

Recall

Recall Number
Z-2142-2020
Event Number
85444
Firm
Inpeco S.A. Via San Gottardo 10 Lugano Switzerland
FEI Number
3010825766
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
April 7, 2020
Terminated
February 6, 2024

Description

Aptio Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Reason

The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.

Action

Urgent: Medical Device Correction notification letters dated 3/30/20 were sent to customers. 1) Visually check that the secondary lane of the Interface Modules listed in Table 1 is empty before sending the Off-line command; or 2) Select the "Going to Off-line" command for the Interface Modules listed in Table 1. This ensures that the Module completes processing samples already inside the Module, releases the tubes and then passes to Off-line status. Your service provider will contact you to schedule the firmware upgrade. Until the service visit please maintain awareness on this notice. Please complete and return the "Customer Letter Receipt Confirmation and Implementation Check form" attached to this letter within 15 days directly to the email address specified in the email communication. Please transfer this notice to whom it might concern. Contact reference person: For any clarification you may need, do not hesitate to contact: Eva Balzarotti - Regulatory Affairs Manager E-mail: [email protected] Phone: (+41) 919118 224

Distribution

US Nationwide including in the states of IL, NY.

Quantity

70 systems