FDA Recall Terminated

ATTUNE CONV RP CR ARTICULATION SURFACE SZ1, 2, 9, and 10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components.

Recall: Z-2140-2015 · Initiated June 12, 2015

Recall

Recall Number
Z-2140-2015
Event Number
71421
Firm
DePuy Orthopaedics, Inc.
FEI Number
1818910
Product Code
MBH
Status
Terminated
Root Cause
Device Design
Initiated
June 12, 2015
Posted
July 20, 2015
Terminated
November 12, 2015
Address
700 Orthopaedic Dr, Warsaw, IN, 46582-3994

Description

ATTUNE CONV RP CR ARTICULATION SURFACE SZ1, 2, 9, and 10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components.

Reason

Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

Action

URGENT INFORMATION: MEDICAL DEVICE CORRECTION was issued for all lots of the affected devices. The device correction is to inform US Distributors and Medical Professionals about the issue, how to prevent the issue, and any actions to take. US Distributors will provide Hospitals/User Facilities with the Medical Professional Notice and Acknowledgment Cards. US Distributors will also work with the sales consultants to ensure all consultants complete sales training regarding this issue.

Distribution

Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and Internationally to CANADA , AUSTRALIA, AUSTRIA, BELGIUM, ENGLAND/ UK, FRANCE, GERMANY, HONG KONG, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JAPAN, MALAYSIA, NETHERLANDS, NEW ZEALAND, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN & IBERIA, SWEDEN, SWITZERLAND, THAILAND, UAE, CHILE, COLOMBIA, MEXICO, PERU, and PUERTO RICO.

Quantity

13964 units