FDA Recall Open, Classified

Blue Silicone, Reusable Silicone Seal 6 - 12 FR. Model Number: CS-B612. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some contain a lumen which allows instruments to be passed through the seal and minimize or prevent fluid leakage during endoscopic procedures.

Recall: Z-2139-2025 · Initiated June 18, 2025

Recall

Recall Number
Z-2139-2025
Event Number
96982
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
HIH
Status
Open, Classified
Root Cause
Packaging change control
Initiated
June 18, 2025
Posted
July 21, 2025
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229

Description

Blue Silicone, Reusable Silicone Seal 6 - 12 FR. Model Number: CS-B612. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some contain a lumen which allows instruments to be passed through the seal and minimize or prevent fluid leakage during endoscopic procedures.

Reason

Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).

Action

Olympus notified consignees on about 06/18/2025 via letter. Consignees were instructed to review the notification and arrange for the return of all affected units on hand, ensure all affected personnel are aware, and notify customers if the affected units were further distributed. Olympus also requested that consignees complete and return the response form acknowledging receipt of the notification.

Distribution

International distribution to the country of Japan.

Quantity

19 units