FDA Recall Open, Classified

Aeos Robotic Digital Microscope, Product Code: PV010

Recall: Z-2139-2024 · Initiated May 13, 2024

Recall

Recall Number
Z-2139-2024
Event Number
94623
Firm
Aesculap Inc
FEI Number
2916714
Product Code
FSO
Status
Open, Classified
Root Cause
Process control
Initiated
May 13, 2024
Posted
June 12, 2024
Address
3773 Corporate Pkwy, Center Valley, PA, 18034-8217

Description

Aeos Robotic Digital Microscope, Product Code: PV010

Reason

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Action

On May 13, 2024, URGENT MEDICAL DEVICE FIELD CORRECTION letters were mailed to customers. Actions to be taken by the customer: 1. Review the Device Correction Notification in its entirety and ensure that all users in your organization of the device and other concerned persons are informed about this voluntary product correction. If you are a distributor, please forward this correction notification to your customers. The correction is to be extended to the end user/customer level. 2. Determine your current inventory of the affected material within your facility, cease use and quarantine product subject to correction. Do not destroy any affected product. (Once your completed Urgent Medical Device Form is received, a representative will be in contact to arrange for inspection and testing). 3. Utilizing the attached "Urgent Medical Device Correction Acknowledgement Form," record the total number of individual units in your posession. If you have no inventory, please enter zero (0) on the form. 4. Return the completed "Urgent Medical Device Correction Acknowledgement Form" to the Aesculap Inc. Quality Assurance Department by faxing the form to (610)849-1197 or e-mail to [email protected] within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). 5. Once we receive your acknowledgement Form, an Aesculap representative will contact you to arrange inspection and testing of your devices by our authorized specialists. Should you identify a problem or safety concern with the product, please report promptly by contacting our Post-market Surveillance Department at 1-833-425-1464. Additionally, adverse reactions of quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting Program wither online, by regular mail or fax.

Distribution

US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.

Quantity

19 units