MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product Usage: The SONATATIi100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired individuals who obtain little or no benefit from acoustic amplification in the best aided condition.
Recall
- Recall Number
- Z-2137-2019
- Event Number
- 81683
- Firm
- MED-EL Elektromedizinische Gereate, Gmbh Furstenweg 77a Innsbruck Austria
- FEI Number
- 3000249315
- Product Code
- PGQ
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- October 16, 2018
- Terminated
- May 22, 2020
Description
MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product Usage: The SONATATIi100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired individuals who obtain little or no benefit from acoustic amplification in the best aided condition.
Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.
Med-EL initially notified customers who had not yet implanted the affected devices on about 10/16/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to not implant the device if still available in inventory. Customers who had implanted the device were notified on about 11/27/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to immediately inform affected clinics of the recall and the long term risk. MED-EL does not recommend a device explantation, as the device may work as intended for years before the electronics may become affected by humidity ingress. Ultimately, accumulation of internal moisture may cause complete failure of the device. Should any uncomfortable or painful sensation be experienced, please stop immediately using the device and contact MED-EL or your local support center for immediate support. MED-EL recommends regular implant checks at least once a year to check the implant status. Additionally the customers were requested to respond to the FSN and confirm receipt.
Worldwide distribution - US Nationwide in the states of California and Illinois. Countries of ARGENTINA, AUSTRALIA, AUSTRIA, BRAZIL, BULGARIA, CHINA, FRANCE, GEORGIA, KOREA, REPUBLIC OF, POLAND, RUSSIAN FEDERATION, SPAIN, SWEDEN.
10 devices