FDA Recall Terminated

Isoloc Software, Version 6.6, 6.5 and 6.0. Part Number MT-NW-425-520. Software computer hard disc reads "ISOLOC 6.6 Copyright NMPE Copyright 1997-2004 NMPE Build #10302006 CIVCO MEDICAL SOLUTIONS Is an EPID-based (electronic portal imaging device) version of localization software for determining the treatment machine position of radiotherapy targets.

Recall: Z-2137-2008 · Initiated June 18, 2008

Recall

Recall Number
Z-2137-2008
Event Number
48463
Firm
Civco Medical Instruments Inc
FEI Number
1937223
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
June 18, 2008
Posted
September 18, 2008
Terminated
March 31, 2011
Address
102 1st St S, Kalona, IA, 52247-9589

Description

Isoloc Software, Version 6.6, 6.5 and 6.0. Part Number MT-NW-425-520. Software computer hard disc reads "ISOLOC 6.6 Copyright NMPE Copyright 1997-2004 NMPE Build #10302006 CIVCO MEDICAL SOLUTIONS Is an EPID-based (electronic portal imaging device) version of localization software for determining the treatment machine position of radiotherapy targets.

Reason

A localization may be produced which has the incorrect moves, and therefore, could place the patient in the incorrect location.

Action

On 06/16/08 Civco Medical Solutions issued an Urgent-Medical Device Correction letter to all its customers by certified mail. This letter states the circumstances in which the device works incorrectly, a temporary work around and the information that the new software version 6.7 will address the issue. The software will be released in August of 2008. Attached to the letter is an ISOLOC Device Correction Notification Customer Feedback form which is to be completed and returned acknowledging the receipt of the Device Correction Notification.

Distribution

Worldwide Distribution --- USA including Puerto Rico and countries of Korea, Portugal, and United Kingdom.

Quantity

235: 163 version 6.6, 60 version 6.5 and 12 version 6.0