FDA Recall Terminated

ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.

Recall: Z-2136-2020 · Initiated January 29, 2020

Recall

Recall Number
Z-2136-2020
Event Number
85467
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JFL
Status
Terminated
Root Cause
Process control
Initiated
January 29, 2020
Posted
May 28, 2020
Terminated
April 28, 2021
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.

Reason

Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.

Action

An Urgent Medical Device Recall letter, dated January 23, 2020 was sent on 01/29/2020 to the affected customers via first class mail and/or email. Customers are advised to: 1) Discontinue use of all ISE Electrolyte Reference Reagent lot M909206 kits. 2) Please review the product Safety Data Sheet (SDS) and discard all ISE Electrolyte Reference Reagent lot M909206 kits according to your laboratorys Standard Operating Procedures and/or according to local regulations. The SDS is available for viewing, printing, and download at https://www.beckmancoulter.com/support/tech-docs. 3) Retrospective review of results is not required. CO2 results are acceptable after a successful calibration and when quality control recoveries are within your laboratorys established QC ranges. 4) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that they are assured customers have received this important communication, they request response within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If customers have any questions regarding this notice, please contact Customer Support Center: " From website: http://www.beckmancoulter.com/customersupport/support " By phone: call 1-800-854-3633 in the United States and Canada. " Outside the United States, contact your local Beckman Coulter Representative. If you need replacement product: " For customers in the United States: o Complete the attached Replacement Order Form and email to [email protected] or fax to (866) 294-7850 OR o Call Client Services at (800) 526-3821 " For customers in Canada: o Complete the attached Replacement Order Form and email to Beckmancoultercanada@be

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the countries of CANADA, JAPAN, LEBANON, SAUDI ARABIA, TAIWAN.

Quantity

2738 vials