FDA Recall Open, Classified

Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Recall: Z-2135-2021 · Initiated June 7, 2021

Recall

Recall Number
Z-2135-2021
Event Number
88158
Firm
Medtronic Vascular, Inc.
FEI Number
3005364322
Product Code
MIH
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
June 7, 2021
Address
3850 Brickway Blvd, Santa Rosa, CA, 95403-8223

Description

Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Reason

During stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover.

Action

On 06/07/2021, the firm initiated a verbal notification regarding an "URGENT MEDICAL DEVICE RECALL" which informed customers that Earlier this week, Medtronic identified the potential for a specific subset of 20Fr Endurant II/IIs Stent Graft Systems to have the potential for the Radiopaque Marker to detach during deployment. There is evidence that devices that used one batch of graft cover subassemblies has a higher potential to exhibit this failure mode. This is an acute failure of the delivery system, there are no concerns with previously implanted stent grafts. We are requesting that you immediately refrain from using this specific subset of products and place it in quarantine until further notice. You will be receiving an official recall notice with instructions for product return for credit or replacement. Your Sales Representative will assist you with this process. We apologize for the inconvenience; however, this action is necessary to ensure the utmost patient safety with the use of our products. On 06/18/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-Day delivery that restated the verbal notification. The Customer Notification Letter instructed customers to: 1)Return all unused affected devices via their sales representative or calling customer service at 1-888-283-7868 or email [email protected]. 2)Complete the enclosed Customer Confirmation Form and email to [email protected]. 3)Forward this notice to all those who need to be aware within their organization. For questions or assistance, contact Customer Service at 1-888-283-7868.

Distribution

U.S.: GA, MA, MD, NY, SC, TX, WA, and WV O.U.S.: Argentina, Australia, Germany, Greece, Israel, Italy, Netherlands, Poland, South Africa, Spain, Switzerland and United Kingdom.

Quantity

18 devices