Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5ix/s5iz Radiology: The Volcano s5/s5i and CORE Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature
Recall
- Recall Number
- Z-2135-2015
- Event Number
- 71654
- Firm
- Volcano Corporation
- FEI Number
- 2939520
- Product Code
- IYO
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- June 22, 2015
- Posted
- July 20, 2015
- Terminated
- April 14, 2017
- Address
- 2870 Kilgore Rd, Rancho Cordova, CA, 95670-6133
Description
Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5ix/s5iz Radiology: The Volcano s5/s5i and CORE Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature
A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.
Urgent Medical Device Correction letters were sent on June 22, 2015 to all affected sites via FED Ex.
Worldwide Distribution-US (nationwide) and regions in Europe and the countries of Japan, Canada, South Africa, Taiwan, Thailand, Singapore, Philippines, Australia, New Zealand, El Salvador, Malaysia, Mexico, Lebanon, Kuwait, South Korea, Kenya, Jordan, Iran, India, Israel, Hong Kong, Egypt, Algeria, Dominican Republic, Costa Rica, Chile, Brazil, Bolivia, and Argentina
A total of 4007 devices, all models