FDA Recall Terminated

Enhancement Medical expression. model # EXP-150. Injectable Gel 1.5cc Syringe with hyaluronic acid concentration of 20 mg/ml. Product Usage: Expression is labelled as an Intranasal Splint, intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. However, Expression is currently being used off-label as a subcutaneously injected dermal filler.

Recall: Z-2135-2014 · Initiated August 7, 2013

Recall

Recall Number
Z-2135-2014
Event Number
68596
Firm
Enhancement Medical, LLC
FEI Number
3008404291
Product Code
LMH
Status
Terminated
Root Cause
Device Design
Initiated
August 7, 2013
Posted
August 8, 2014
Terminated
September 29, 2017
Address
10201 Innovation Drive, Suite 450, Wauwatosa, WI, 53226

Description

Enhancement Medical expression. model # EXP-150. Injectable Gel 1.5cc Syringe with hyaluronic acid concentration of 20 mg/ml. Product Usage: Expression is labelled as an Intranasal Splint, intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. However, Expression is currently being used off-label as a subcutaneously injected dermal filler.

Reason

Manufacturer made a change in the production process that resulted in a change in final gel weight. RECALL EXPANDED 7/8/2014 Firm expanded their recall to include all lots of product.

Action

Enhancement Medical, LLC called all Independent Sales Representatives and then provided each representative a list of consignees with impacted product and instructed the Representatives to disclose low gel weight issue and potential for common treatment responses. Representatives were instructed to exchange the affected product. EXPANDED Enhancement Medical sent an Urgent Medical Device Recall letter beginning July 8, 2014 to all consignees whom received product. The letter identified all lots as being recalled, the reason for the recall, recommended that Expression not be injected subcutaneously, and asked for product to no longer be used and be returned to the recalling firm. An acknowledgement Form was asked to be mailed, e-mailed or faxed back to the firm. Customers with questions were instructed to call 414-918-4280.

Distribution

Worldwide Distribution - USA (nationwide including Puerto Rico) and Internationally to Canada.

Quantity

17,875