Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-36-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 36mm, Novation GXL Liner, 10 Deg Face, 36mm. (note: inadvertently left out of original posting) Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-40-XX, Novation GXL Liner, 10 Deg Face, 20mm ID. (note: inadvertently left out of original posting) Orthopedic hip implant component.
Recall
- Recall Number
- Z-2132-2021
- Event Number
- 88126
- Firm
- Exactech, Inc.
- FEI Number
- 1038671
- Product Code
- LZO
- Status
- Open, Classified
- Root Cause
- Unknown/Undetermined by firm
- Initiated
- June 29, 2021
- Posted
- July 22, 2021
- Address
- 2320 Nw 66th Ct, Gainesville, FL, 32653-1630
Description
Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-32-XX, Novation GXL Liner, 10 Deg Face, 32mm. Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-36-XX, Acumatch GXL 15 Degree, +5 Lat Liner, 36mm, Novation GXL Liner, 10 Deg Face, 36mm. (note: inadvertently left out of original posting) Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 138-40-XX, Novation GXL Liner, 10 Deg Face, 20mm ID. (note: inadvertently left out of original posting) Orthopedic hip implant component.
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
The firm issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 6/28/2021 via email on 6/29/2021 that were addressed to the attention of Exactech Agents, Representatives, and Distributors in possession of affected products.
Distribution US nationwide.
89,050 total