GE Senographe Pristina - Product Usage: Senographe Pristina generates digital mammographic images that can be used for screening and in the diagnosis of breast cancer. Senographe Pristina is intended to be used in the same clinical applications as traditional film-based mammographic systems.
Recall
- Recall Number
- Z-2131-2019
- Event Number
- 83096
- Firm
- GE Healthcare, LLC
- FEI Number
- 2126677
- Product Code
- MUE
- Status
- Terminated
- Root Cause
- Equipment maintenance
- Initiated
- May 16, 2019
- Terminated
- December 16, 2020
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Senographe Pristina - Product Usage: Senographe Pristina generates digital mammographic images that can be used for screening and in the diagnosis of breast cancer. Senographe Pristina is intended to be used in the same clinical applications as traditional film-based mammographic systems.
Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure.
On May 17, 2019, GE Healthcare sent letters to all of their consignees advising them of the recall, providing them with new instructions for use, and requesting that they complete the enclosed MEDICAL DEVICE NOTIFICATION ACKNOWLEDGEMENT RESPONSE REQUIRED forms. GE will follow-up and correct these systems,
Worldwide distribution - US nationwide distribution in the states of AZ, CA, FL, GA, IL, MI, MN, NC, NJ, NY, PA, TX, and VT, and countries of Algeria, Australia, Belgium, Canada, Chile, Columbia, Czech Republic, Finland, France, Germany, Hungary, Iceland, Japan, Korea (Republic of), Kuwait, Lebanon, New Zealand, Norway, Poland, Saudi Arabia, Senegal, South Africa, Sri Lanka, Sweden, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
82