FDA Recall Open, Classified

Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D00U003; 2. Percepta Serena Solara application, Software Model Number D00U004; 3. Cobalt Crome application, Software Model Number D00U005; 4. Micra VR application, Software Model Number D00U006; 5. Micra AV application, Software Model Number D00U007; 6. Viva Consulta Syncra Advisa Ensura application, Software Model Number D00U008; 7. Claria Amplia Compia application, Software Model Number D00U009; 8. Evera MRI application, Software Model Number D00U010; 9. Visia AF application, Software Model Number D00U011; 10. Viva Brava Evera application, Software Model Number D00U012; 11. Micra VR2 AV2 application, Software Model Number D00U022.

Recall: Z-2127-2025 · Initiated June 17, 2025

Recall

Recall Number
Z-2127-2025
Event Number
97125
Firm
Medtronic, Inc.
FEI Number
2182208
Product Code
KRG
Status
Open, Classified
Root Cause
Software design
Initiated
June 17, 2025
Posted
July 17, 2025
Address
8200 Coral Sea St Ne, Mounds View, MN, 55112-4391

Description

Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D00U003; 2. Percepta Serena Solara application, Software Model Number D00U004; 3. Cobalt Crome application, Software Model Number D00U005; 4. Micra VR application, Software Model Number D00U006; 5. Micra AV application, Software Model Number D00U007; 6. Viva Consulta Syncra Advisa Ensura application, Software Model Number D00U008; 7. Claria Amplia Compia application, Software Model Number D00U009; 8. Evera MRI application, Software Model Number D00U010; 9. Visia AF application, Software Model Number D00U011; 10. Viva Brava Evera application, Software Model Number D00U012; 11. Micra VR2 AV2 application, Software Model Number D00U022.

Reason

A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025.

Action

Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTIFICATION to its consignees on 06/18/2025 via UPS 2nd day delivery. The notice announced that a software update was available for the Medtronic CareLink SmartSync Device Manager (SmartSync), explained the issue which the update was correcting, risk, and requested the following: Customer Actions: " Update SmartSync (to version 4.2.3 or higher). Refer to release notes provided by your local Medtronic representative for update instructions. If needed, Medtronic representatives are available to work with you to install or update the SmartSync application(s) on your tablet. See Appendix for instructions on how to verify the software update is complete. " Sign and return the enclosed confirmation form and share this notice with those who need to be aware within your organization or with any organization where SmartSync may be in use. Questions regarding this information should be directed to the local Medtronic Representative or Medtronic Programmer Technical Services at 800-638-1991, Option 1.

Distribution

Worldwide

Quantity

3909 units