PE CONNECTOR 3/8 STRAIGHT 20/BX, Sterile, Single use, Product Usage: Catheter Connector/Accessory
Recall
- Recall Number
- Z-2127-2017
- Event Number
- 77081
- Firm
- Teleflex Medical
- FEI Number
- 3005747797
- Product Code
- GCD
- Status
- Terminated
- Root Cause
- Labeling False and Misleading
- Initiated
- April 21, 2017
- Terminated
- January 27, 2020
- Address
- 2917 Weck Dr, Research Triangle Park, NC, 27709-0186
Description
PE CONNECTOR 3/8 STRAIGHT 20/BX, Sterile, Single use, Product Usage: Catheter Connector/Accessory
Labeling: The expiration date is not stated on the labeling, there is a potential for expired product to be used.
Consignees were notified via Urgent Medical Device Recall letter on April 21, 2017. The letter identified the affected product, problem and actions to be taken. They were instructed to discontinue use and quarantine any products with the affected catalog numbers and lot numbers. They were to return the product, complete the enclosed Recall Acknowledgement Form and fax it to the Attn: Customer Service or email to [email protected]. A Recall Acknowledgment Form was enclosed to be completed and returned via fax even if they have no stock.
Worldwide Distribution - U.S. Nationwide. Foreign consignees: Australia, Canada, China, Columbia, India, San Salvador, Singapore and South Korea.
19,043 units in total