FDA Recall Terminated

Microscan plus Negative Breakpoint Combo 4 Sold outside of US only. Product Usage: MicroScan Synergies plus Panels & MicroScan Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.

Recall: Z-2126-2013 · Initiated August 21, 2013

Recall

Recall Number
Z-2126-2013
Event Number
66058
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2919016
Product Code
LON
Status
Terminated
Root Cause
Process control
Initiated
August 21, 2013
Posted
September 11, 2013
Terminated
May 15, 2014
Address
2040 Enterprise Blvd, West Sacramento, CA, 95691-3427

Description

Microscan plus Negative Breakpoint Combo 4 Sold outside of US only. Product Usage: MicroScan Synergies plus Panels & MicroScan Rapids plus (Negative) - For use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification for gram-negative bacteria.

Reason

Confirmed increase in false susceptible and false intermediate misreads for Imipenem and/or meropenum on Synergies plus Negative and rap ID/S plus Negative panels read on Walk Away System.

Action

Siemens sent an Urgent Field Safety Notice letters dated August 2013 to all US customers via over night delivery, receipt requested. Letters for foreign customers are being translated and will be sent as soon as possible. The letter identified the affected product, reason for field correction, risk to health problem and actions to be taken by customer. Customers were instructed to suppress all reporting of sensitive and intermediate results for imipenem and meropenem. Siemens also recommends that the customer consider the need to review previous test results, conduct patient follow-up, and/or repeat testing. Customer were requested to complete the attached Field Correction Effectiveness Check form and fax it to 302-631-8467. For questions contact your Siemens Technical Support Provider or Distributor.

Distribution

Worldwide Distribution - USA Nationwide in the states of: AK, SC, IA, MO GA, WI, TX, KS, CA, IL, CO, AZ, SD, IN, MD, FL, MN and the countries of:Japan, France, Algeria, Germany, Austria, Romania, Russian Federation, South Africa, Kazakhstan, Belgium, Great Britain, Turkey, Spain, Italy, Saudi Arabia, Qutar, Poland, Iraq, Unite Arab Emrites, Algeria, Oman, Greece, Portugal, and Ireland.

Quantity

57,600 panels in total