IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only.
Recall
- Recall Number
- Z-2124-2012
- Event Number
- 62555
- Firm
- Iris Diagnostics
- FEI Number
- 2023446
- Product Code
- JJW
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- July 9, 2012
- Posted
- August 1, 2012
- Terminated
- January 27, 2014
- Address
- 9172 Eton Ave, Chatsworth, CA, 91311-5805
Description
IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only.
The firm recalled after it received complaints of IRISpec CA/CB two part Urine Chemistry Control, 101131-12, where the nitrite results read negative for the part B control, which should be positive.
The firm initiated the recall on 07/09/12 by sending out a Medical Device Recall notification to customers via e-mail. The recall notification identified the affected product, reason for recall, and instructions on the recall. Customers were to confirm how many sets of unopened or partially opened IRISpec CA/CB that was in their inventory, dispose of the affected product, and provide written confirmation of the product's disposition. The letter also stated that Iris Diagnostics would replace any unused inventory that was returned at no charge to the customer. Contact information for the recall was also provided. The communications also included a fax response sheet for return/replacement.
Nationwide Distribution
862