FDA Recall Open, Classified

VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);

Recall: Z-2122-2025 · Initiated June 10, 2025

Recall

Recall Number
Z-2122-2025
Event Number
97047
Firm
Ortho-Clinical Diagnostics, Inc.
FEI Number
1000136573
Product Code
DIH
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
June 10, 2025
Posted
July 17, 2025
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);

Reason

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Action

On June 10, 2025, "IMPORTANT PRODUCT CORRECTION NOTIFICATION" letters were sent to distributors and customers. CUSTOMER LETTERS: REQUIRED ACTIONS " Monitor VITROS CHE Slides QC results to detect a drift in VITROS Performance Verifiers I and II below 2x Within-lab SD limits. If QC results are below 2x WLSD, perform QC again using freshly reconstituted vials of VITROS Performance Verifiers I and II. " Complete the enclosed Confirmation of Receipt form no later than June 18, 2025. " Save this notification with your User Documentation or post this notification near your laboratory's fluid reconstitution location until the issue has been resolved. " Please forward this notification if the affected products were distributed outside of your facility. " If your laboratory has experienced the issue described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization. QuidelOrtho's investigation is ongoing, and we are working to identify root cause. We will communicate again after root cause has been determined and the issue has been resolved. Please note that until this issue is resolved, future lots of VITROS Performance Verifiers I and II may not meet the 7-day reconstituted stability for CHE. DISTRIBUTORS: REQUIRED ACTIONS " Please send the enclosed customer letter and customer Confirmation of Receipt form to all customers who have been shipped VITROS Chemistry Products Performance Verifiers AND VITROS CHE Slides from August 1, 2023, to date. " Complete the enclosed distributor Confirmation of Receipt form no later than June 18, 2025. " Please forward this notification if the affected products were distributed outside of your facility. If you have further questions, please contact the Global Services Organization at 1-800-421-3311.

Distribution

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.

Quantity

27, 330 US/ 19,679,730 OUS