FDA Recall Terminated

Thyroid Collar, Buckle Closure, Product Number TS4, Part Numbers RBTS4-R, RBTS4-T2T, RETS4-R, RLTS4-R, RLTS4-SEP, RLTS4-T2T, SBTS4-R, SETS4-T2T, SLTS4-R, UETS4-R Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Recall: Z-2122-2019 · Initiated June 7, 2019

Recall

Recall Number
Z-2122-2019
Event Number
83144
Firm
Burlington Medical, LLC
FEI Number
1000120804
Product Code
EAJ
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
June 7, 2019
Posted
July 30, 2019
Terminated
April 28, 2023
Address
3 Elmhurst St PO Box 3194, Newport News, VA, 23603-1137

Description

Thyroid Collar, Buckle Closure, Product Number TS4, Part Numbers RBTS4-R, RBTS4-T2T, RETS4-R, RLTS4-R, RLTS4-SEP, RLTS4-T2T, SBTS4-R, SETS4-T2T, SLTS4-R, UETS4-R Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Reason

The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.

Action

Urgent Medical Device Recall notification letters were sent to customers on 6/7/19. Please take the following actions: Cease using recalled products immediately, quarantine the products, and prepare the products for return shipment to Burlington Medical. Contact Burlington Medical at (800) 221-3466 for specific return instructions. Note that Burlington Medical Customer Care will work with you to facilitate the return of recalled products, including by issuing you a Recall Tracking number. Burlington Medical will replace all recalled products at no cost to you. The replacement product will be the same size and model as the recalled product; if the same size and model is not available or has been discontinued, then Burlington Medical will replace the recalled product with a Burlington Medical product that is similar to the recalled product. Our Customer Care team will arrange for free shipment of the replacement product. If you have any questions, please call your Burlington Medical representative or Burlington Medical Customer Care at (800) 221-3466. The recall was expanded an additional Urgent Medical Device Recall notification letters were sent to customers on 7/3/19.

Distribution

US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.

Quantity

807