FDA Recall Open, Classified

Brand Name: GMAX Product Name: SYR 10ML Control syringe Model/Catalog Number: TS3210L-M Product Description: NON-Sterile syringes without needles for single use Component: No

Recall: Z-2121-2024 · Initiated April 5, 2024

Recall

Recall Number
Z-2121-2024
Event Number
94520
Firm
Jiangsu Shenli Medical Production Co., Ltd. No.
FEI Number
3003902757
Product Code
FMF
Status
Open, Classified
Root Cause
No Marketing Application
Initiated
April 5, 2024
Posted
May 20, 2024
Address
20 Changzheng Road, Zhengluzhen Wujin Changzhou China

Description

Brand Name: GMAX Product Name: SYR 10ML Control syringe Model/Catalog Number: TS3210L-M Product Description: NON-Sterile syringes without needles for single use Component: No

Reason

Piston syringes sizes and configurations are out of the range of devices cleared under the firm's 510(k).

Action

On 05/14/2024, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter informing customers that Jiangsu Shenli Medical Production Co., Ltd. is recalling piston syringes that were distributed without premarket clearance. The Recalling Firm stated that they received an FDA Warning Letter that the piston syringes were not legally marketed devices. Customers are instructed to: Examine their inventory and stop/cease use of all syringe products subject to this action and report the inventory quantity to Jiangsu Shenli Medical Production Co., Ltd. Considering the complicated return process, and high cost of returning the products, the firm is instructing customers to dispose of these recalled syringes in the U.S, using their facility s approved/established procedures for disposal of medical products, or by contacting a local third party company to assist with the disposal of the recalled syringes. Please report the destruction method and cost with Shenli before final implementation. After product disposal, please provide Shenli with confirmation of destruction of products. For questions or assistance - contact the general manager at email [email protected] or telephone number 0086-519-88731071

Distribution

U.S. Nationwide distribution in the states of CA, FL, GA, IL, TN, and VA. After customers place US orders, then arrange production, then deliver to the cargos to customer designated port or warehouse, mostly in Shanghai port. The shipping company will ship the cargos to different destination such as Chicago, Los Angeles, Long beach, Miami, Savannah, Nashville, and Norfolk.

Quantity

348800