Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.
Recall
- Recall Number
- Z-2121-2017
- Event Number
- 77083
- Firm
- Medtest Holdings, Inc.
- FEI Number
- 1827821
- Product Code
- JJY
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- May 15, 2014
- Posted
- May 20, 2017
- Terminated
- June 2, 2017
- Address
- 5449 Research Dr, Canton, MI, 48188-2261
Description
Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.
The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.
Customers were notified on 05/5/14by sending a technical bulletin (via email, fax, US Mail and posting on website) to all customers and distributors that had received lot 303104 of C7590, lot 303105 of C7591 or C7592 kits with either lot since product was approved to ship. All product shipped after 05/15/2014 included the updated package inserts. Any questions concerning the implementation of the new controls ranges may be directed to Pointe Scientific's Technical Service Department at 1-800-757-5313. Thank you for your continued support of Pointe Scientific.
Nationwide
132919 Vol vials